by ShaggyDog18@gmail.com, JULY 2020
github: https://github.com/ShaggyDog18/SD_MHZ19B
Arduino library for MH-Z19B NDIR CO2 Module by Zhengzhou Winsen Electronics Technology Co., Ltd
License: GNU GPLv3
MH-Z19B NDIR infrared gas module is a common type, small size sensor, using non-dispersive infrared (NDIR) principle to detect the existence of CO2 in the air, with good selectivity, non-oxygen dependent and long life. Built-in temperature compensation; and it has UART output and PWM output.
This compact library features access to all module features though an abstracted class and methods to manage the module implemented in user-level functionality.
The library does not work with PWM module signal. Communicates with the module by Hardware or Software UART.
CO2 Measurement Range: 0-2000ppm or 0-5000ppm. Accuracy: ±(50ppm+3%).
SD_MHZ19B( Stream& serial )
- Class Constructor~SD_MHZ19B()
- Class Destructoruint16_t getPPM(void)
- reads data from the module; returns value of CO2 concentration in ppm; Data are verified and validated by calculating checkSum of the data received. RetunrsFALSE
i.e. ZERO value if communication or CRC error occurs.int8_t getTemp(void)
- returns module temperature in degrees Celcium. Should be called after getPPM() function; otherwise returns a previous value. Rather inaccurate ±2*C; is used for internal compensation.void setAutoCalibration( bool _autoCalib )
- toggles Auto Calibration (Automatic Baseline Correction - ABC) ON/OFF. Settrue
for Enabled/ON;false
for Disabled/OFF (default). Refer to Auto Calibration Notes below.bool getABCstatus(void)
- gets Auto Calibration ABC status (undocumented feature). Returnstrue
for Enabled/ON,false
- Disabled/OFF.void calibrateZeroPoint(void)
- calibrates Zero point. During zero point calibration, the sensor must work in stable gas environment (400ppm) for over 20 minutes.bool calibrateSpanPoint( uint16_t _spanPoint )
- performsSpan
calibration. Do aZeroPoint
calibration beforeSpan
calibration. Make sure the sensor worked under a certain level co2 for over 20 minutes. Suggest using 2000ppm as span, at least 1000ppm. Default value is 2000ppm. Returns true if the requested Span value is OK.bool setDetectionRange( uint16_t _detectionRange )
- sets detection range, default is 0~2000ppm. The range could be 2,000-5,000ppm. Returnstrue
if the requested range value is OK.uint8_t getStatus(void)
- reads module status (undocumented feature), Returns 0 if OK; The value is available after successful getPPM() reading only; i.e. to be called after getPPM(). Otherwise, returns a previous value.
For more details on the library use refer to the example that utilizes major library methods.
if Auto Calibration (Automatic Baseline Correction - ABC) is ON, the sensor itself performs a zero point judgment and automatic calibration procedure intelligently after a continuous operation period. The automatic calibration cycle is every 24 hours after powered on. The zero point for automatic calibration is 400ppm.
This function is usually suitable for indoor air quality monitor such as offices and homes, not suitable for greenhouse, farm and refrigeratory where this function should be off.
CO2 Level, ppm | Description |
---|---|
250 - 400 |
Normal background concentration in outdoor ambient air |
400 - 1,000 |
Concentrations typical of occupied indoor spaces with good air exchange |
1,000-2,000 |
Complaints of drowsiness and poor air. |
2,000-5,000 |
Headaches, sleepiness and stagnant, stale, stuffy air. Poor concentration, loss of attention, increased heart rate... |
The library developed for Arduino UNO, NANO, Pro Mini, ESP8266, etc.