- 2x Ø1" Polaris K1, Polaris K1H mirror mounts with installed mirrors
- 2x 5.5 mm stilts with a M4 thru hole through the middle of it's body
- 2x M4x25 mm capscrews
- 2x M4 nuts
- 2x ThorLabs part # SD1 M4 to M6 adapter
- 2x ThorLabs part : RC1 rail carrier
- 1x needle nosed pliers
- 1x 3 mm hex key/driver
Put the M4x25 mm capscrew through a SD1 M4 to M6 adapter, then the capscrew/adapter assembly through the bottom side of the RC1 rail carrier. The 5.5 mm stilt then goes over and through the M4 thru hole, so that there is about 3-5 mm of threading of the M4 capscrew poking out. Set the base of the mirror on the stilt through the hole that will put the horizontal adjustment screw on the left side of the assembly, and screw a M4 nut over the remaining threading of the M4 capscrew. Grab and hold the M4 nut with the needle nosed pliers and from the bottom of the RC1 rail carrier use the 3 mm hex key/driver to screw the capscrew tight. Further angular adjustments will need to be made to the position of these assemblies when aligning the laser.
Note: The video shows the use of Ø1/2" mirror mounts, the Madison system uses them but the Dresden system uses the Ø1" mirror mounts. The video was made in Madison during the construction of it's OpenSPIM
{% include video id="g3BHLcUeKzA" width="400" height="243" caption="Assembly small mirror mount to rail carrier" %}
- 6x Ø1/2" optical mounts with installed optical components (2x f< = 25 mm achromat, 2x f = 50 mm achromat, and 2x 1 mm ground glass alignment disk)
- 6x 14.95 mm stilts with a M4 thru hole through the middle of it's body
- 6x M4x25 mm capscrews
- 6x ThorLabs part # SD1 - M4 to M6 adapter
- 6x ThorLabs part # RC1 - rail carrier
- 1x 3 mm hex key/driver
Put the M4x25 mm capscrew through a SD1 M4 to M6 adaptor, then the capscrew/adaptor assembly through the bottom side of the RC1 rail carrier. The 14.95 mm stilt then goes over and through the M4 thru hole, so that there is about 5 mm of threading of the M4 capscrew poking out. On the bottom/base of the lens mount is a threaded hole for M4 capscrew, rotate the mount around until it is snug then give it a 1/4 turn to tighten it with the 3 mm hex key/driver. Further angular adjustments will need to be made to the position of these assemblies when aligning the laser.
{% include video id="Amb_-zSIRYc" width="400" height="243" caption="Assembly spherical lens mount to rail carrier" %}
- 1x VA100/M Vertical slit mechanism
- 1x 13.17 mm stilt with a M4 thru hole through the middle of it's body
- 1x M4 x 22 mm capscrew (cut a few millimeters of threading off of a 25 mm capscrew)
- 1x ThorLabs part # SD1 M4 to M6 adapter
- 1x ThorLabs part # RC1 rail carrier
- 1x 3 mm hex key/driver
Put the M4x22 mm capscrew through a SD1 M4 to M6 adapter, then the capscrew/adapter assembly through the bottom side of the RC1 rail carrier. The 13 mm stilt then goes over and through the M4 thru hole, so that there is about 2 mm of threading of the M4 capscrew poking out. On the bottom/base of the vertical slit mechanism is a threaded hole for M4 capscrew, rotate the mount around until it is snug then give it a 1/4 turn to tighten it with the 3 mm hex key/driver. Further angular adjustments will need to be made to the position of this assembly when aligning the laser.
{% include video id="i-MgtiCzTiQ" width="400" height="243" caption="Assembly vertical slit mount to rail carrier" %}
Go back to Step by step assembly